ContaminationSensor CS 1000 CS1000 hydraulic installation
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 23/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
Flow rate, differential pressure
p
and viscosity
characteristics
Differential pressure
p
and viscosity
characteristics. All the values indicated in
the figures below apply regardless whether the direction of flow is A->B or B->A.
Note that the permissible measured volumetric flow is 30 … 500 ml/min.
If you are unable to achieve the required flow values, we offer an extensive line of
accessories with various conditioning modules.
For example:
You are using a fluid with a viscosity
of 46 mm²/s at a pressure difference
p
of
~0.9 bar, so that you achieve a flow rate of approx. 100 ml/min.
The flow rate depends on the viscosity of the medium and the differential pressure
p
via the sensor.